Technical Data:
Thermal Focus1.0 x 10 mm Nominal voltage60 kV, max.
Line focus (viewed at 6° takeoff angle)0.1 x 10 mm Filament voltage9 V
Point focus (viewed at 6° takeoff angle)1.0 x 1.0 mm Filament current3.6 A, max.

Tube Power:
for Mo, W2400 W
for Cu, Ag2000 W
for Co, Cr1800 W
for Fe1500 W
 
Cooling Requirements:
Min. Water Flow3.5 l/min
Max. Water Temp.35° C
Max. Water Pressure0.8 MPa

Inherent Filtration:
300 µm Be Window


Ordering Information:
Part Number Description Part Number Description
XRD 60 10 CUS XRD 60-10 Cu S
Copper target, short anode
XRD 60 10 CUL XRD 60-10 Cu L
Copper target, long anode
XRD 60 10 WS XRD 60-10 W S
Tungsten target, short anode
XRD 60 10 WL XRD 60-10 W L
Tungsten target, long anode
XRD 60 10 MOS XRD 60-10 Mo S
Moly target, short anode
XRD 60 10 MOL XRD 60-10 Mo L
Moly target, Long anode
XRD 60 10 CRS XRD 60-10 Cr S
Chrome target, short anode
XRD 60 10 CRL XRD 60-10 Cr L
Chrome target, Long anode
XRD 60 10 COS XRD 60-10 Co S
Cobalt target, short anode
XRD 60 10 COL XRD 60-10 Co L
Cobalt target, Long anode
XRD 60 10 FES XRD 60-10 Fe S
Iron target, short anode
XRD 60 10 FEL XRD 60-10 Fe L
Iron target, Long anode
XRD 60 10 AGS XRD 60-10 Ag S
Silver target, short anode
XRD 60 10 AGL XRD 60-10 Ag L
Silver target, Long anode




Dimension Short Anode Long Anode
A17.3 mm30.1 mm
B21 mm34 mm
C105 mm118 mm
D194 mm207 mm
E213 mm226 mm

100% Replacement Warranty
In the unlikely event that an XRD 1000 tube should fail due to material or workmanship within the first twelve months or 2500 hours of operation, whichever occurs first, the XRD tube will be replaced.